Proswift Reverse Phase Specifications
 
Printer Friendly

Proswift Reverse Phase Specifications

Reversed-Phase
Column name
PepSwiftProSwift
RP-1S
ProSwift
RP-2H
ProSwift
RP-3U
Column dimensions
(i.d. x length)
 
100 µm x 5 cm
200 µm x 5 cm
500 µm x 5 cm
4.6 × 50 mm4.6 × 50 mm4.6 × 50 mm
Base matrix material
Polystyrene-divinylbenzenePolystyrene-divinylbenzenePolystyrene-divinylbenzenePolystyrene-divinylbenzene
Surface chemistry
PhenylPhenylPhenylPhenyl
Protein binding capacity per mL of polymer
1.9 mg/mL5.5 mg/mL1.0 mg/mL0.5 mg/mL
AngiotensinInsulinLysozymeLysozyme
Bed height
50 mm46 mm44 mm40 mm
Column volume
0.4 µL
1.6 µL
9.8 µL
0.76 mL0.73 mL0.65 mL
pH Range
2.5–81–141–141–14
Recommended flow rate
0.8–1.0 µL/min
2.5–3.0 µL/min
15–20 µL/min
0.5–4 mL/min1–10 mL/min1–16 mL/min
Maximum flow rate
1.2 µL/min
5.0 µL/min
30 µL/min
5 mL/min14 mL/min20 mL/min
Maximum pressure
5000 psi2800 psi2800 psi2800 psi
34.4 Mpa19.3 Mpa19.3 Mpa19.3 Mpa
Operating pressure
<2900 psi<1500 psi<1500 psi<1500 psi
20 Mpa10.3 Mpa10.3 Mpa10.3 Mpa
Temperature
70 °C70 °C70 °C70 °C
Solvent compatability
Most common organic solventsMost common organic solventsMost common organic solventsMost common organic solvents

   
 Proswift Reverse Phase Specifications 
 
Order Literature   Privacy   Feedback   Terms of Use   Trademarks   Site Map