| | | Photodiode Array Detector |
| The Photodiode Array Detector is the detector of choice for laboratories who need maximum flexibility. It can be used both for routine applications and applications such as method development and peak characterization. The 1 nm spectral resolution allows you to distinguish between closely related UV-VIS spectra and the comprehensive set of 3-D data tools in the Chromeleon® Chromatography Management Software makes it easy to process and report your data. - Low noise (<±10 µAU) and drift (<1 mAU/h) guarantee high sensitivity
- Stainless steel and PEEK flow cells provide maximum application flexibility.
- Data collection rates of up to 20 Hz allow detection and characterization of very sharp peaks
- Front access to prealigned cells and lamps simplifies routine maintenance and therefore minimizes instrument downtime
Performance Examples
| | | Photodiode Array Detector Specifications | | Wavelength Range | 190–800 nm | | Noise | <±1.0 × 10 –5 AU, at 254 nm, 2-s rise time, flowing water <±1.5 × 10 –5 AU, at 520 nm, 2-s rise time, flowing water | | Drift | <1 × 10 –3 AU/h (typical <5 × 10 –4 AU/h) | | Wavelength Accuracy | ±1.0 nm, self-calibration with D2 lines, verification with built-in holmium oxide filter | | Spectral Resolution | 1 nm | | Linearity | Less than 5% RSD up to 2.0 AU Less than 3% RSD up to 1.5 AU | | Principle | Single-beam principle with simultaneous measurement across the photodiode array. | | Light Source | Tungsten and deuterium lamp | | Optical Bench | 1024 element photodiode array bench without moving parts | | Pixel resolution | 0.7 nm | | PC Connection | USB | | Signal Output | 5 detection channels plus 3-D field digitally transmitted | | Analog Outputs | Four; 0–3 AU, selectable, 1000 mV range | | Flow Cell | PEEK or SST, fused silica windows | | Cell Path Length | Analytical: 10 mm Semimicro: 9 mm Semipreparative: 0.4 mm | | Cell Volume | Analytical: 13 µL Semi-micro: 3.1 µL | | Pressure | PEEK <300 psi (2.1 MPa) SST <500 psi (3.5 MPa) | | Power Requirements | 90–265 V ac, 47–63 Hz (autosensing, no adjustment needed) | | Operating Temperature Range | 4–40 °C (40–104 °F). Operation at less than 10 °C assumes constant operation | | Operating Humidity Range | 5–95% relative, noncondensing | | Dimensions (h x w x d) | 19 x 42 x 50 cm (7.5 x 16.5 x 19.7 inch) | | Weight | 18.3 kg (40 lb) |
| | Order Information | | PDA-3000 UltiMate 3000 Photo Diode Array Detector | 5080.0020 | | Analytical Flow Cell PDA-3000, PEEK | 6080.0200 | | Analytical Flow Cell PDA-3000, SST | 6080.0210 | | Semi-Micro Flow Cell PDA-3000, PEEK | 6080.0220 | | Semi-Micro Flow Cell PDA-3000, SST | 6080.0230 | | Semi-Prep Flow Cell PDA-3000, PEEK | 6080.0240 |
| |
|
|  |  | UltiMate LC and LC-MS Detectors | | Choose from a wide range of sensitive detectors. Operate even the MSQ™ Plus Mass Spectrometer, together with the other instruments in your system, from a single Chromeleon® control panel. |
| Modules | | | UltiMate ® 3000 LC modules achieve new levels of flexibility, performance, reliability, and ease of use. |
| LC and LC/MS Overview | Discover the UltiMate ® 3000 LCi series, and learn how LCi solutions can improve performance and productivity in your laboratory. |
| LCi Solutions |  | Find out how the UltiMate® 3000 LCi Series can provide you with LC solutions for your chromatographic challenges. |
|
|