| AS Autosampler |
 |
|
The Dionex ICS-3000TM AS Autosampler provides flexibility, high precision, reliability, ruggedness, and ease of use. Automation has never been
more powerful.
-
Free up your schedule and lab time with automated sample handling for your IC.
-
Improve productivity and throughput with innovative, simultaneous, and sequential sampling modes.
-
Improve your results with automated sample preparation.
-
Eliminate complex laboratory procedures with automated standard preparation.
Key Features Provide Performance and Versatility
-
Perform full loop, partial loop, limited sample, sample preparation, sample dilution, pre-column derivatization, matrix elimination,
sample pre-concentration, and more.
-
Consistent operation at RSDs of less than 0.3%
-
Specially designed 10-mL polystyrene sample vials with wide openings that are easy to fill, easy to handle, and permit large-volume
injections
-
Well Plate ready for limited sample and high through-put applciations.
-
Easy access to components for effective troubleshooting
-
All-PEEK flow paths that ensure compatibility with aqueous and reversed-phase eluents and protect sensitive samples from metallic
contamination
-
Control through Chromeleon®, or module front panel
-
Moving-needle design to guarantee reliable sampling from a variety of vial sizes
-
Easy syringe priming
-
Sample Prep option that saves time and labor by automating your sample and standard preparations
- Sample tray temperature control option for thermally sensitive samples that offers precise, reliable control over a temperature
range of 4–60 °C
Learn how the AS autosampler will improve the productivity of your laboratory in "Techniques to Increase Sample Throughput
and Productivity", in the November 2005 issue of LCGC The Peak magazine.
|
|
|
| AS Autosampler Specifications |
| Sample capacity |
10 mL vials - 49 Samples
1.5 and 2.0 mL vials - 99 Samples
Well Plates (tray holds two well plates) - 96 Samples per plate for a total of 192 Samples |
| Vial size |
10-mL polystyrene vials
1.5-mL glass vials
0.3 mL and 1.5 mL Polypropylene vials
0.45, 0.5 and 2.0 mL Well Plates |
| Number of injections per vial |
1–99 |
| Minimum sample volume |
10 µL can be sampled from a 300-µL microvial
20 µL can be sampled from 500 µL in a 10-mL vial |
| Variable volume range |
1–100 µL in 0.1-µL increments
100–1000 µL in 1-µL increments |
| Injection valve |
PEEK Rheodyne with Tefzel® rotor seal
|
| Injection loop size |
25-µL standard; other sizes available |
| Injection precision |
Fixed loop <0.3% RSD at 20 µL or greater; variable <0.5% RSD at 20 µL |
| Dilution precision |
<1.0% RSD for a 1:100 dilution |
| Dispensing precision |
<0.2% RSD by weighing |
| Carryover |
<0.01% with 500-µL flush volume |
| Ambient temperature range |
10–40 °C |
| Humidity range |
10–90% relative humidity, noncondensing |
| Sample Modes |
Full-loop, Partial loop, Limited Sample partial loop, Simultaneous Injection, Sequential, Sequential Concentrate |
|
| Order Information |
| AS Autosampler, Base Configuration |
063102 |
| AS Autosampler, with Sample Preparation Option |
063103 |
| AS Autosampler, with Sample Tray Temperature Control Option |
063106 |
| AS Autosampler, with Sample Prep and Tray Temp Control Options |
063107 |
| AS Autosampler, with Injection Valve |
061786 |
| AS Autosampler, with Injection Valve and Sample Tray Temperature Control |
061788 |
| AS Autosampler, Simultaneous Injection Configuration without Valves |
063104 |
| AS Autosampler, Simultaneous Injection with Sample Prep Option |
063105 |
| AS Autosampler, Simultaneous Injection Configuration with Internal Valves |
063101 |
| AS Autosampler, Simultaneous Injection Configuration with Internal Valves and Sample Prep Option |
063100 |
|
|
|
|
|
|
|
|
© 2008 Dionex Corporation. All rights reserved. www.dionex.com
|
|