ICS-2000 Specifications
 
Printer Friendly

ICS-2000 Specifications

ICS-2000 Integrated Reagent-Free IC System Specifications

Analytical Pump and Fluidics

Type

Serial dual-reciprocating pistons, microprocessor-controlled constant stroke, variable speed

Construction

Chemically inert, metal-free PEEK pump heads and flow paths compatible with aqueous eluents of pH 0–14 and reversed-phase solvents

Pump Operating Pressure

0–35 MPa (0–5000 psi)

Flow Rate Range

0.00–5.00 mL/min in 0.01 increments without changing pump heads

Flow Precision

<0.1%

Flow Accuracy

<0.1%

Pressure Ripple

<1%

Eluent On-Off Valve

Standard

Piston Seal Wash

Dual-pump head, wash can be continuous when connected to rinse solution supply

Pressure Alarm Limits

Upper limit 0–35 MPa or 0–5000 psi in one unit (MPa or psi) increments; lower limit can be set up to one unit lower than upper limit

Vacuum Degas

Yes, optional, automatic control

Eluent Bottles

Polypropylene

Eluent Bottle Pressure

Not required

Injection Valve

6-port, 2-position Rheodyne valve, electrically activated

Maximum Column Lengths

250 mm analytical column with 50 mm guard column

Column Heater

Standard

Operating Temperature Range

30 to 60 °C (86 to 140 °F); minimum
5 °C above the ambient temperature. Settable range is equal to working range.

Temperature Stability

±0.5 °C at sensor

Temperature Accuracy

±0.5 °C at sensor; at 40 °C

Eluent Generation

Standard

Eluent Types

KOH, LiOH, NaOH
K2CO3
K2CO3/KHCO3
MSA

Gradient Profiles

Option; Combination of unlimited number of linear, convex and concave positive and negative gradient profiles

Concentration Increments

0.01 mM

Concentration Range

0.1–100 mM (Depending on eluent used)

Flow Rate

0.1–3.0 mL/min

Maximum Operating Pressure

21 MPa (3000 psi)

Maximum Solvent Concentration

Anions: 25% methanol
Cations: no solvents

Eluent Regeneration Support

Yes, with optional RFIC-ER Controller

Eluents

Carbonate and carbonate/bicarbonate combinations up to 20 mM

MSA up to 34 mM

Flow Rates

0.01–2.00 mL/min

Continuous Operation with 4 L of eluents

Up to 28 days or 2000 samples, typically

Always On, Always Ready Capable

Yes, standard feature

Remains Fully Calibrated for Extended Periods
(<28 days)

Yes, standard feature. Results are traceable to a single calibration

System Wellness

Consumables usage monitoring for predictive maintenance

Maximum Operating Pressure

21 MPa (3000 psi)

Operating Temperature Range

4–40 °C

Suppressors and Control

Chemical Suppression

2 mm and 4 mm anion and cation, membrane suppression bed type

Displacement Chemical Regeneration

2 mm and 4 mm anion and cation, membrane suppression bed type

Electrolytic Suppression- Self Regenerating.

2 mm and 4 mm anion and cation. Both membrane and MonoDisc suppression bed types available

Electrolytic Suppression- Self Regenerating with External Water Mode

2 mm and 4 mm anion and cation. Both membrane and MonoDisc suppression bed types available

Current Control Range

SRS, 0–500 mA in 1 mA increments
AES, 0–50 mA in 1 mA increments

Salt Converter

Available in 2 and 4 mm versions

AMMS-ICE

Available in 2 and 4 mm versions

Carbonic Acid Removal for Anions

Yes: ASRS 300 plus CRD-300

Non-Suppressed

Yes, supported

Suppressor Wear Parts

None: peristaltic pump and inline filters are not required

Suppression Capacity

Anion SRS 300 (4 mm): 200 µeq/min
Cation SRS 300 (4 mm): 110 µeq/min
Anion SRS 300 (2 mm): 50 µeq/min
Cation SRS 300 (2 mm): 37.5 µeq/min
Anion MMS 300 (4 mm): 150 µeq/min
Cation MMS 300 (4 mm): 150 µeq/min
Anion MMS 300 (2 mm): 37.5 µeq/min
Cation MMS 300 (2 mm): 37.5 µeq/min
Anion AES: 25 µeq/min
Cation AES:25 µeq/min
AMMS-ICE: N/A

Suppressor Void Volumes

SRS 300 (4 mm): <50 µL
SRS 300 (2 mm): <15 µL
MMS 300 (4 mm): <50 µL
MMS 300 (2 mm): <15 µL
AMMS-ICE 300 (4 mm): <50 µL
AMMS-ICE 300 (2 mm): <15 µL
AES (Anion or Cation): <35 µL

Conductivity Detector Electronics and Flow Cell

Type

Microprocessor-controlled digital signal processor

Cell Drive

8 kHz square wave

Linearity

1%

Resolution

0.00238 nS/cm

Full-Scale Output Ranges

Digital signal range 0–15,000 µS
Analog signal range 0–15,000 µS

Electronic Noise

±0.1 nS/cm when background conductivity is 0–150 µS/cm
±2 nS/cm when background conductivity is 151–3200 µS/cm

Filter

Rise times from 0 to 10 s, user selectable

Temperature Compensation

Fixed at 1.7%/°C at cell temperature

Temperature Range

Ambient +7 °C; 30 to 55 °C

Temperature Stability

<0.01 °C

Cell Electrodes

Passivated 316 stainless steel. Compatible with MSA

Cell Body

Chemically inert polymeric material

Cell Volume

<1 µL

Heat Exchanger

Inert, tortuous-path for low axial dispersion

Maximum Cell Operating Pressure

10 MPa (1500 psi)

Autosampler

Automation Using Autosampler

Dionex AS40, AS, AS-HV, or third-party Autosamplers

Sequential/Simultaneous Injection

Yes, depending on autosampler capabilities

Automated Dilution

Yes, available with AS Autosampler

Dilution Factor, AS Autosampler

1:1 to 1:1000

Dilution Time, AS Autosampler

15 seconds with sample overlap

Inline Sample Degassing

Yes, optional with CRD 300/200

Inline Filtration

Yes, AS40 Autosampler or inline filter

Automation Flexibility

High, conditionals using Chromeleon and post run features

System Software and Control

Software

Chromeleon Chromatography Management Software, supports Microsoft Windows® 2000, XP, or Vista

Automated Procedure Wizards

Yes, standard feature

System Smart Startup and Shutdown

Yes, standard feature

System Wellness and Predictive Performance

Yes, standard feature

Data Trending Plots

Yes, numerical device parameters can be plotted

Virtual Column Simulator

Yes; evaluation mode standard,
isocratic and gradient optional

Application Templates

Yes, standard feature

Multivendor Automation Support of 3rd Party

Yes, fully controls over 300 different instruments from more than 30 manufacturers, including GC, HPLC, and MS

3-D Software for PDA and MS

Yes, optional

Customizable System Control Panels

Yes, standard feature

System Status Virtual Channels

Yes, standard feature

Power Failure Protection

Yes, standard feature

Sequential Injection

Yes, standard feature for independent dual system analysis

System Trigger Commands and Conditionals

Yes, standard feature

Daily Audit Trail

Yes, standard feature

Sample Audit Trail

Yes, standard feature

Multiple Network Control and Network Failure Protection

Yes, optional

System Calibration Storage

Yes, factory, present and previous. Completely user selectable

Customized Reporting

Yes, standard feature with unlimited report workbooks

Control Modes

Full control through front panel and Chromeleon software; alternative control through TTL or relay closures; two relay outputs, two TTL outputs, four programmable inputs

Semi Automated System Qualification

Yes, automated IQ/OQ/PQ procedures produce detailed, comprehensive qualification reports using minimal operator time

USB Communication Protocol

One input; one built-in two-port USB hub

Physical Specifications

Dimensions (h × w × d)

56.1 × 22.4 × 53.3 cm (22.1 × 8.8 × 21 in.)

Weight

24.5 kg (54 lb)

Power Requirements

100–240 V ac, 50–60 Hz, autoranging

Operating Temperature

4–40 °C (40–104 °F); cold-room-compatible (4 °C) as long as system power remains on

Operating Humidity Range

5–95% relative, non-condensing

Leak Detection

Built-in, optical sensor

ICS-2000 Ion Chromatography System Data Sheet   219K
ICS-2000/1500/1000 Brochure   2MB
     
  ICS-2000 Specifications  
 
Order Literature   Privacy   Feedback   Terms of Use   Trademarks   Site Map