The Dionex SE 400 and SE 500 Solvent Evaporator systems are designed to rapidly evaporate organic solvents or water from ASE® collection vials or bottles. The systems are equipped with heated blocks and flow manifolds that accelerate the evaporation
process. The solvent evaporator systems decrease sample preparation times and increase laboratory productivity. The Dionex
Solvent Evaporator Systems feature:
-
Positive-pressure heated nitrogen flow
-
Heated top manifold and bottom blocks
-
Heated block vortex action (SE 500)
-
Compatible with ASE 200, 40- or 60-mL vials
-
Compatible with ASE 100 and 300, 250-mL collection bottles
-
Interchangeable manifolds and blocks to accommodate different size collection vessels
-
Solvent level can be checked without interrupting the evaporation process
Application Areas for the solvent evaporators include:
| SOLVENT EVAPORATION TIMES FOR SE 500 |
| Solvent |
Conditions (PSI ) |
Starting Volume (mL) |
Time to 1 mL (min) |
| *acetone/DCM |
40 |
50 |
26 |
| *acetone/hexane |
40 |
50 |
19 |
| **water |
40 |
50 |
120 |
| *acetone/DCM |
20 |
50 |
33 |
| *acetone/hexane |
20 |
50 |
21 |
| **water |
20 |
50 |
175 |
|
*Temperature of vortex block: 50 °C. Temperature of the manifold: 70 °C. Speed 800 rpm. Pulse 70 per min, duty: 90% **Temperature of vortex block: 90 °C. Temperature of the manifold: 80 °C. Speed 800 rpm. Pulse 70 per min, duty: 90%
|
| Part numbers |
|
SE 500-24 Holds 24 40 or 60 mL vials, 120 V
|
063221
|
|
SE 500-24 holds 24 40 or 60 mL vials, 240 V
|
063218
|
|
SE 500-6 holds six 250 mL bottles, 120 V
|
063220
|
|
SE 500-6 holds six 250 mL bottles, 240 V
|
063219
|
|
SE 400-8 holds eight 40 or 60 mL vials, 120 V
|
063222
|
|
SE 400-2 holds two 250 mL bottles, 120 V
|
063224
|
|
SE 400-8 holds eight 40or 60 mL vials, 240 V
|
063223
|
|
SE 400-2 holds two 250 mL bottles, 240 V
|
063225
|
|
SE 500 heated manifold 250 mL bottles, 120 V
|
063231
|
|
SE 500 heated manifold 250 mL bottles, 1240 V
|
063230
|
|
SE 500 heated manifold 40 or 60 mL vials, 120 V
|
063233
|
|
SE 500 heated manifold 40 or 60 mL vials, 240 V
|
063232
|
|
250 mL bottle heating block for SE 500
|
063235
|
|
40/60 mL vial heating block for SE 500
|
063234
|
|
40/60 mL vial block for SE 400
|
063237
|
|
250 mL bottle block for SE 400
|
063236
|
|
40/60 mL vial polypropylene manifold for SE 400
|
063239
|
|
250 mL bottle polypropylene manifold for SE 400
|
063238
|
|
Evaporation needles 2 in
|
063229
|
|
Evaporator Plug
|
063228
|
|
Solvent Cabinet for SE 500
|
063227
|
|
Solvent Recovery system SE 500 with solvent cabinet
|
063226
|
|
| SE 400 and SE 500 Solvent Evaporation Systems Specifications |
| Power Requirements: |
SE 400: 5.0 A, 120 or 240 V, 50/60 Hz SE 500: 10.0 A, 120 or 240 V, 50/60 Hz |
| Operating Environment: |
SE 400: 0–90% relative humidity, 32–100 °F; 0–37 °C SE 500: 0–90% relative humidity, 32–100 °F; 0–37 °C |
| Gas Requirements: |
SE 400: 20–40 psi analytical grade SE 500: 20–40 psi analytical grade |
| Dimensions: |
SE 400: 38.1 × 19.0 × 15.2 cm (15 × 7.5 × 6 in) h × w × d
SE 500: 54.0 x 48.9 x 58.4 cm (21.25 × 19.25 × 23 in, with cabinet) h × w × d Platform size: 29.5 × 36.2 cm ( 11.625 × 14.25 in) |
| Weight: |
SE 400: 8 lb (3.6 kg) SE 500: 75 lb (34.0 kg) |
|