IonPac AS10 Anion-Exchange Column

The IonPac AS10 is a high-capacity, hydroxide-selective, anion-exchange column designed for the isocratic and gradient separation of inorganic anions and organic acids. 

  • Use the high-capacity AS10 column for the analysis of inorganic anions in high nitrate samples.
  • Simplified Reagent-FreeTM IC operation is provided by the EG40 or EG50 Eluent Generator, or the RFC-30 Controller, which requires only a deionized water source to produce potassium hydroxide eluent.

The column provides excellent resolution of weakly retained low-molecular-weight aliphatic acids. The selectivity and high capacity of the AS10 permits the analysis of trace-level analytes in high-ionic-strength matrices. The AS10 resin strongly retains nitrate, which makes this column well suited for the analysis of trace anions in nitric acid and nitrate salts. Use with the EG40 or EG50 Eluent Generator for simplified eluent preparation for applications using less than 100 mN hydroxide. Use the Anion Self-Regenerating Suppressor (ASRS® ULTRA) with the AS10 column.

IonPac AS10 Anion-Exchange Column  Specifications
Mobil phase compatibility

pH 0–14; 0–100% of HPLC solvents

Substrate characteristics

Bead Diameter: 8.5 µm

Cross-Linking (%DVB): 55%

Dimensions

IonPac AS10 Analytical Column:
2 × 250 mm and 4 × 250 mm

IonPac AG10 Guard Column:
2 × 50 mm and 4 × 50 mm

Latex characteristics

Functional Group: Alkanol quaternary ammonium

Latex Diameter: 65 nm

Latex Cross-Linking: 5%

Capacity

42.5 µeq (2 × 250 mm analytical column)

170 µeq (4 × 250 mm analytical column)

Order Information
IonPac AS10 Analytical Column (4 x 250 mm) 043118
IonPac AG10 Guard Column (4 x 50 mm) 043119
IonPac AS10 Analytical Column (2 x 250 mm) 043123
IonPac AG10 Guard Column (2 x 50 mm) 043124
Data Sheets
IonPac AS10 Anion-Exchange Column and the IonPac AC10 Concentrator Column Data Sheet   197K
IonPac AS10 Analytical Column Manual   1MB
AN 78: Determination of Trace Anions in Concentrated Hydrofluoric Acid   84K
AN 85: Determination of Trace Anions in Organic Solvents   149K
TN 46: Determination of Trace Anions in Concentrated Glycolic Acid   192K
 
 
© 2008 Dionex Corporation. All rights reserved. www.dionex.com