Columns and Accessories > IC and RFIC Columns > Hydroxide-Selective Anion Columns > IonPac AS21 Anion-Exchange Column
 
Chromatograms /  Specifications  /  Order Info  /  Data SheetsPrinter Friendly
IonPac AS21 Anion-Exchange Column

The IonPac AS21 hydroxide selective anion-exchange column is specifically designed for trace perchlorate in drinking water prior to MS/MS detection.  The capacity and selectivity allows for the fast analysis of perchlorate in drinking water at the low µg/L concentration.  The AS21 column is the specified column in U.S. EPA Method 331.0.

  • Recommended anion-exchange column for fast separation of perchlorate prior to MS/MS detection.
  • Optimized for methylamine or hydroxide mobile phases.
  • Specified column for U.S. EPA Method 331.0.
  • Optimum capacity: 45 µeq per column. (2 × 250-mm column).
  • Operate at ambient or elevated temperatures. Column selectivity is optimized for a 30 °C operating temperature to ensure reproducible retention times in all environmental conditions.
  • Compatible with HPLC organic solvents that optimize the performance of the electrospray interface (ESI) to a MS or MS/MS detector, modify column selectivity, or for effective column clean up.

The IonPac AS21 hydroxide-selective anion-exchange column is designed for the fast separation of perchlorate in drinking water, ground water, and other matrices prior to detection with MS/MS. The AS21 2 × 250 mm column format was specifically developed with MS/MS compatibility to allow use of volatile mobile phases such as methylamine. The IonPac AS21 is the specified column in U.S. EPA Method 331.0.

IonPac AS21 Anion-Exchange Column  Specifications
Dimensions:

IonPac AS21 Analytical Column:
2 × 250 mm

IonPac AG21 Guard Column:
2 × 50 mm

Maximum Operating Pressure:

3000 psi

Mobile Phase Compatibility:

pH 0–14; 0–100% HPLC solvents

Substrate Characteristics:

Bead Diameter and Pore Size:
7 µm (2 × 250 mm); super macroporous resin (2000 Å)
11 µm (2 × 50 mm); microporous resin (<1 Å)

Crosslinking (%DVB): 55%

Ion Exchange Group:

Functional Group: Alkanol quaternary ammonium ion

Functional Group Characteristics:

Ultralow Hydrophobicity

Capacity:

45 µeq (2 × 250-mm analytical column)

1.5 µeq* (2 × 50-mm guard column)

*Guards are packed with a low-capacity microporous resin.

Column Construction:

PEEK with 10–32 threaded ferrule style end fittings. All components are nonmetallic.

Order Information
IonPac AS21 Analytical Column (2 x 250 mm)063009
IonPac AG21 Guard Column (2 x 50 mm)063071
Data Sheets
IonPac AS21 Anion-Exchange Column Data Sheet   1MB
IonPac AS21 Analytical Column Manual   602K

MSQ Plus Mass Spectrometric Detector

The MSQTM  Plus is the smallest and most sensitive single quadrupole MS detector.

   
  Columns and Accessories > IC and RFIC Columns > Hydroxide-Selective Anion Columns > IonPac AS21 Anion-Exchange Column 
 
Order Literature   Privacy   Feedback   Terms of Use   Trademarks   Site Map
© 2008 Dionex Corporation. All rights reserved. www.dionex.com