Columns and Accessories > IC and RFIC Columns > Cation Columns > IonPac CS10 Cation-Exchange Column
 
Chromatograms /  Specifications  /  Order Info  /  Data SheetsPrinter Friendly
IonPac CS10 Cation-Exchange Column

The IonPac® CS10 is a moderate-capacity, sulfonate-functionalized, cation-exchange column designed for the isocratic separation of inorganic cations using hydrochloric acid or diaminopropionic acid eluents. The CS10 can also be used for biogenic amines. The CS12A column is the recommended column for the fast analysis of inorganic cations. The new CS17 column is the column of choice for biogenic amines.

  • The CS10 can be used for analysis of the common inorganic cations.
  • The CS10 column can be used for biogenic amines.
  • The CS12A column provides improved performance for fast analysis of inorganic cations and ammonium.
  • The new CS17 column is recommended for hydrophobic and polyvalent amines.
  • Sample matrices include environmental waters, power plant waters treated with ammonium, morpholine or ethanolamine, chemical additives, chemical process solutions, scrubber solutions, plating baths, and industrial solvents.

The IonPac CS10 is designed for the isocratic separation of lithium, sodium, ammonium, potassium, magnesium, and calcium using hydrochloric acid or diaminopropionic acid eluents with suppressed conductivity detection. The common inorganic cations can be determined in diverse sample matrices, including drinking water, power plant waters, soil extracts, acid digests, chemical additives, chemical process solutions, scrubber solutions, and plating baths. Use the Cation Self-Regenerating Suppressor (CSRS® ULTRA) with the CS10 column.

IonPac CS10 Cation-Exchange Column  Specifications
Dimensions

IonPac CS10 Analytical Column:
4 × 250 mm

IonPac CG10 Guard Column:
4 × 50 mm

Maximum operating pressure

4000 psi (27.5 MPa)

Mobile phase compatibility

pH 0–14; 0–100% HPLC solvents (i.e., methanol, acetonitrile).

Resin characteristics

8.5 µm ethylvinylbenzene/divinylbenzene substrate agglomerated with 200-nm MicroBeadTM cation-exchange latex

Latex characteristics

Ion-Exchange Group: Sulfonic acid

Latex Cross-Linking: 5%

Latex Diameter: 200 nm

Hyrophobicity: Medium

Capacity

80 µeq (4 × 250 mm analytical column)

Column construction

PEEK with 10-32 threaded ferrule-style end fittings. All components are nonmetallic.

Order Information
IonPac CS10 Analytical Column (4 x 250 mm)043015
IonPac CG10 Guard Column (4 x 50 mm)043016
Data Sheets
IonPac CS10 Cation Exchange Column Data Sheet   238K
IonPac CS10 and CS11 Analytical Columns Manual   386K
IonPac CS10 Cation Exchange Column   165K
CS12A Column

The IonPac® CS12A is the recommended column for the fast separation of mono- and divalent cations.

CS17 Column
The IonPac® CS17 is recommended column for hydrophobic and polyvalent amines, including biogenic amines amd diamines.
SRS 300

The Self-Regenerating Suppressor, (SRS®  300), is used for electrolytically regenerated suppression of ion chromatography applications requiring high capacity, fast start-up, and sensitivity.

   
  Columns and Accessories > IC and RFIC Columns > Cation Columns > IonPac CS10 Cation-Exchange Column 
 
Order Literature   Privacy   Feedback   Terms of Use   Trademarks   Site Map