The ICS-3000 ED Electrochemical Detector features a new cell design along with a newly designed reference electrode. The cell
is more rugged due to the new handle design which provides consistent torque to the working electrode. Exciting innovations
will create new opportunities for your laboratory.

ED Features
-
One-piece reference electrode provides consistency and reliability
-
Handle design provides consistent torque to cell electrode for consistent installation and mounting of working electrode
-
Uses microprocessor-controlled digital signal processing
-
Supports DC amperometry, pulsed amperometry, or integrated amperometry detection modes
-
Detection capabilities include use of multiple waveforms and multiple integration times to optimize detection conditions for
individual analytes
-
Three-dimensional display of the raw integrated amperometry data set similar to photodiode array data display, with cross-hairs
on an isoamperometric plot used to select “slices” of the plot along the applied voltage axis (to render a chromatogram) and
along the time axis (to render a voltammogram)
-
Three-dimensional wireframe rendering, printing, color selection, display of apex, and other spectra on peaks in chromatograms
are echoed in the integrated amperometry implementation
-
The integrated amperometry mode provides complete freedom to change the waveform profile's number of segments, duration of
each segment, and voltage applied at each segment
-
Control through Chromeleon® software or locally through TTL inputs
-
Mounts within the DC compartment in either of two locations
-
No tools are required for installation
-
Cell and detector electronics integrated to minimize noise, maximize electrical isolation and shielding, and maximize thermal
stability
-
Innovative, built-in electronics for easy calibrations and diagnostics
-
Dual-detection configurations (detectors in series or parallel)
|
| ICS-3000 ED Electrochemical Detector Specifications |
| Electronics Type |
Microprocessor controlled digital signal processing |
| Electronic Noise |
IPAD < 80 pC DC Amperometry < 5 pA |
| Potential Range |
-2.0–2.0 V in 0.001 V increments |
| Signal Range: Digital and Analog |
DC amperometry: 5 pA to 74 µA Integrated amperometry: 50 pC to 200 µC |
| Filter |
0–10s response time |
| Control Mode |
Local or remote using relay closures, TTL, of Chromeleon using DC module |
| Cell Body |
Titanium body and inlet tubing Compatible with 0.5 and 4 mm id Pressure rating is 120 psi when fully assembled |
| Working Electrodes |
Gold, glassy carbon, platinum, and silver |
| Reference Electrode |
pH-Ag/AgCl combination |
| Autoranging |
Autoranging digital amperometry with Chromeleon |
| Analog Output |
User selectable full scale of 10, 100, or 1000 mV |
| Working Volume |
< 0.5 µL |
| Maximum Operating Pressure |
100 psi (0.7 MPa) |
| Dimensions (h x w x d) |
2.7 × 6.5 × 3.9 in (6.9 × 16.7 × 9.9 cm) |
| Weight |
400 g (1.6 lb.) |
|