Rapid Separation Thermostatted Column Compartment
 
Specifications  /  Order Info  /  Data SheetsPrinter Friendly
Rapid Separation Thermostatted Column Compartment

The Rapid Separation Thermostatted Column Compartment is an essential component of the UltiMate® 3000 Rapid Separation LC System, and can be operated precisely and accurately at temperatures up to 110 °C. The benefit of high temperatures is that they reduce system backpressure and typically provide sharper peaks. With efficient precolumn heating and postcolumn cooling technology, the TCC-3000RS is perfectly equipped to ensure the best chromatography under these challenging conditions. This high-end column compartment is highly flexible, too, supporting up to six LC columns with a length of up to 30 cm, and can easily be upgraded with up to two integrated column switching valves.

The TCC RS provides column ID chips for automated documentation in audit trails and reports. In addition, automated diagnostic tests and on-line troubleshooting guide minimize instrument downtime and facilitate regulatory compliance.
The column compartment is designed to meet the requirements of any laboratory interested in ultrafast chromatography with even the most challenging thermostatting requirements.

  • Temperature range of 5-110 °C, with an accuracy of ± 0.5 °C and a precision of ± 0.1 °C.
  • Precise temperature control, even under varying ambient conditions. [more]
  • Fast heat-up/cool down times: typically 12 min from 20 °C to 50 °C; typically 15 min from 50 °C to 20 °C.
  • Active postcolumn eluent cooler for ensuring lowest detector noise and drift even at extreme column temperatures and when using fast gradients.
  • Provides access to the x2 dual technology . Take advantage of techniques such as automated application switching  and increase your laboratory performance and productivity even further.
Performance Examples

Rapid Separation Thermostatted Column Compartment  Specifications
Temperature range

5 °C to 110 °C (maximum 18 °C below ambient)

Temperature accuracy

±0.5 °C

Temperature stability

+/-0.1 °C

Temperature precision

±0.1 °C

Column capacity

6 columns, maximum 30 cm

Heatup/Cooldown time

Typically 12 min from 20 °C to 50 °C
Typically 15 min from 50 °C to 20 °C

Switching valves

Up to two 2-position, 6 or 10-port valves

Post column eluent cooler

Active temperature control

Dimensions (h x w x d)

18 x 42 x 51 cm (7.1 x 16.5 x 20 in.)

Power requirements

100 - 120 V, 60 Hz
200 - 240 V, 50 Hz

PC connection

USB

Inputs/Outputs

2 digital inputs, 2 relay outputs

GLP

Dionex AutoQ routines for instrument qualification
Column identification for up to 4 columns

Order Information
TCC-3000RS Rapid Separation Thermostatted Column Compartment 5730.0000
Data Sheets
No datasheet available at this time.
RSLC - Dionex Solution for UHPLC

Experience ultrahigh performance chromatography while maintaining reliability and ease-of-use.

Rapid Separation LC
The UltiMate Rapid Separation LC (RSLC) system offers a unique combination of high flow rate, high pressure, and ultrafast data collection.
LC and LC/MS Overview

Discover the UltiMate ®  3000 LCi series, and learn how LCi solutions can improve performance and productivity in your laboratory.

Modules
UltiMate ® 3000 LC modules achieve new levels of flexibility, performance, reliability, and ease of use.
Chromeleon Chromatography Management System
Whether you use HPLC, IC, or GC—whether your scope is a single instrument, a laboratory, a department, or an enterprise—Chromeleon adapts to your needs. 
     
  Rapid Separation Thermostatted Column Compartment  
 
Order Literature   Privacy   Feedback   Terms of Use   Trademarks   Site Map
© 2008 Dionex Corporation. All rights reserved. www.dionex.com